Product | Aplo Shell 400 | Aplo Shell 700 |
Features | High rigidity, high flow rate, large sample loading capacity | High rigidity, high flow rate, large sample loading capacity |
Application | Separation and purification of biological macromolecules such as viruses, exosomes and vaccines | Separation and purification of biological macromolecules such as viruses, exosomes and vaccines |
Matrix | High-rigidity agarose | |
Particle size | 70-120 μm | 65-85 μm |
Functional group | Octylamine | |
Shell exclusion molecular weight | 400 kDa | 700 kDa |
Flow velocity | <600 cm/h | <600 cm/h |
Max. pressure | <0.5 MPa | |
pH stability | 3-12(operational), 2-14(CIP, short term) | |
Chemical stability | Stable in commonly used buffers, such as 1 M NaCl, 1 M NaOH, 6 M guanidine hydrochloride, 8 M urea, 70% ethanol, 30% isopropyl alcohol, etc. | |
Storage conditions | 20% ethanol; 4-30℃ | |
Pack size | Cat. No. | |
Aplo Shell 400 | Aplo Shell 700 | |
1ml XTrap Prepacked columns | HC5080111 | HC5080211 |
5ml XTrap Prepacked columns | HC5080112 | HC5080212 |
5ml XScreen Prepacked columns | HC5080113 | HC5080213 |
10ml | HC5080100 | HC5080200 |
25ml | HC5080101 | HC5080201 |
100ml | HC5080102 | HC5080202 |
500ml | HC5080103 | HC5080203 |
1L | HC5080104 | HC5080204 |
5L | HC5080105 | HC5080205 |
10L | HC5080106 | HC5080206 |
1.1 Packing Buffer Preparation
Purified water; degas byultrasonication for 15 min.
1.4 Column Packing (example: 16 mm column diameter, 10 cm bedheight)
Mix the resin slurry thoroughly and pour it slowly into the columnalong a glass rod. If the liquid level remains below the top of the column, addpacking buffer to fill the column. Connect the adaptor to the chromatographysystem, start the pump, and remove air from the tubing and top screen at anappropriate flow rate; then pause the system. Insert the adaptor into thecolumn at a 45° angle, secure it, and tighten the sealing ring while preventingair from entering. Remove the bottom plug and direct the bottom tubing into awaste container. Pack at 60 cm/h until the resin-bed interface stabilizes.Increase the linear velocity to 600 cm/h, with the high-pressure alarm set to0.5 MPa, and maintain the final rate for 45 min. Mark the resin-bed interfaceand stop the system. Reinstall the bottom plug and disconnect the top of thecolumn from the pump. Loosen the adaptor sealing ring slightly, lower theadaptor to 2 mm below the resin surface, and retighten the seal. Connect thetop and bottom column tubing to the chromatography system and perform acolumn-efficiency test.[TR1]
Table 2. Column EfficiencyTest Methods
Note*: CV (Column Volume) refers tocolumn volume and is abbreviated as CV below.
Calculate N/m and As from the UVcurve (or conductivity curve) using the following equations:
Figure 1. SchematicDiagram for Column Efficiency Calculation
Connect thechromatography column to the purification system. Equilibrate the column withat least 3-5 CV of equilibration buffer until the pH and conductivity of thecolumn effluent match those of the binding buffer. After equilibration, zerothe UV detector.[TR2]
Wash the column with2-3 CV of 1 M NaOH containing 27% propanol. Equilibrate with at least 3-5 CV ofequilibration buffer until the effluent pH and conductivity match those of thebinding buffer.[TR3]
If the column will be usedimmediately after cleaning, wash with 3-10 CV of binding buffer.
5.1 Wash the column with 3-5 CV ofbinding buffer.
5.2 Wash the column with 2-5 CV of0.5 M NaOH.
5.3 Soak the column in 0.5-1 M NaOHfor 1 h.
5.4 Complete sanitization by washingthe column with 5-10 CV of binding buffer at pH 7-8.
Store packed columns sealed andimmersed in 20% ethanol to prevent microbial contamination.
Note: Before using stored resin,perform CIP and equilibrate with at least 5 CV of binding buffer.
Keep the residence time constant tomaintain stable dynamic binding capacity.
Ensure uniform sample concentrationand consistent elution conditions.
Table 3. TroubleshootingRecommendations